7. Maintenance, Cleaning, Diagnosis
This chapter describes the user maintenance and cleaning procedures for the Programat P710. Only those tasks are listed
that may be performed by dental professionals. All other tasks must be performed by qualified service personnel at a
certified Ivoclar Vivadent Service Centre.
7.1 Monitoring and maintenance
The time for these maintenance procedures depends on the frequency of use and the working habits of the users. For that
reason, the recommended times are only approximations.
Hazardous area
Part
When
Check all plug-in connections for correct fit.
Various external connections
weekly
Check if the furnace head opens smoothly and without excessive
noise.
Furnace head opening mechanism monthly
Check if the thermocouple is straight and in the right place
Thermocouple
weekly
Check the insulation for cracks and damages. If the insulation is
worn down it has to be replaced by a certified Ivoclar Vivadent
After Sales Service Centre. Fine hairline cracks on the surface of the
insulation are harmless and do not adversely affect the function of
the furnace.
Insulation
monthly
Check if the sealing rims of the furnace head and furnace base are
clean and undamaged.
Sealing rims of the furnace head
and the furnace base
weekly
Check the membrane-sealed keypad for visible damage. If the
membrane-sealed keypad is damaged, it has to be replaced by a
certified Ivoclar After Sales Service Centre.
Membrane-sealed keypad
weekly
Conduct temperature check:
Use the temperature checking set to check and adjust the
temperature in the furnace.
Firing chamber
every six
months
Check the quartz glass cylinder to make sure the quartz glass is not
defective.
Firing chamber
daily
Check if there is condensate in the vacuum hose or firing chamber.
Vacuum hose,
Firing chamber
monthly
This furnace has been developed for typical use in dental laboratories. If the product is used in a production
enterprise, for industrial application and for continuous use, premature ageing of the expendable parts has to be
expected. The expendable parts are, e.g. the heating muffle, furnace head seal or the insulation material.
Expandable parts are not covered by the warranty. Please also observe the shorter service and maintenance intervals.
In general, the furnace head should not be replaced since the components (furnace head and furnace base) have
been coordinated with each other. However, if the furnace head must be replaced for maintenance reasons, sub-
sequent temperature calibration is required.