Ivoclar Vivadent EP5000-G2 User Manual
Page 15

15
Step 3:
Placing the firing plate for the investment
ring (5)
The firing plate for the investment ring (5) can
now be placed on the firing plate holder (48).
Ensure that the furnace head mounting mark (41)
is aligned with the furnace base mounting mark
(42).
Make sure that the firing plate
holder (48) is not damaged by
mounting the furnace head.
Step 4:
Connections
Connect the cables of the furnace head with the
furnace base. Proceed as follows:
– Insert the thermocouple plug (26) (make sure
that the polarity of the plug is correct)
– Insert the heater plug (28)
– Insert the press drive plug (60)
5
48
28
27
26
60
41
42
50
54