beautypg.com

Product specifications – Ivoclar Vivadent Lumamat 100 User Manual

Page 19

background image

19

9. Product Specifications

Lumamat 100

This chapter contains all the
relevant product specification.

9.1 Delivery form

1 Lumamat 100
1 Power cord
1 Extra fuse
1 Extra lamp
1 Screw driver
1 Operating Instructions
1 Warranty card
1 Standard accessories set
1 Test set

Colours:
Standard colour:
White

RAL 9016

Special colours:
Salmon

RAL 3014

Aquamarine

RAL 5014

Turquoise

RAL 6027

9.2 Technical data

Power supply
Single-phase alternating current:
220–230 V / 50–60 Hz
100 V / 50–60 Hz
110 V / 50–60 Hz
120 V / 50–60 Hz

Tolerated voltage fluctuations:
+/– 10 %

Power consumption:
max. 750 W

Type of lamp:
Fluorescent lamps (only use
Ivoclar Vivadent lamps)

Electrical fuses:
Value for all voltage versions:
For 220–230 V:

T 5A

Has to comply with IEC 127
For 100–120 V:

T 8A

Has to be UL and CSA listed.

Dimensions of fuses:
Diameter 5 x 20 mm

Dimensions:
Width:

323 mm

Depth:

440 mm

Height:
closed

272 mm

open

540 mm

Lightwave range:
400–580 nm

Weight:
13.4 Kg

Safety information

This apparatus has been
constructed according to the
following directives:
– Directive 73/336/EEC, low-

voltage guidelines

– Directive 89/336/EEC, radio

protection /electromagnetic
compatibility

Furthermore, the following
standards were observed:
– IEC 61010-1
– UL 3101-1
– CSA 1010.1

9.3 Acceptable operating

conditions

Acceptable temperature range
5–40 °C / 41–104 °F

Acceptable humidity range
Maximum relative humidity at
31 °C (87 °F), gradually
decreasing to 50 % at 35 °C
(95 °F), condensation excluded.

Acceptable ambient pressure
The unit has been tested for use
at altitudes of up to 2000 m
above sea level.

9.4 Acceptable

transportation and
storage conditions

Acceptable temperature range
-30 – +80 °C (-22 – +176 °F)

Acceptable humidity range
Maximum relative humidity at
31 °C (87 °F) gradually decrea-
sing to 50 % at 40 °C (104 °F),
condensation excluded.

Acceptable ambient pressure
500-1060 mbar.

Use only original packaging with