Maintenance, cleaning, and diagnosis – Ivoclar Vivadent Lumamat 100 User Manual
Page 15

15
7. Maintenance, Cleaning, and Diagnosis
Item
Frequency
Cleaning material/measure
Dust filter
Monthly
Batting or replacing
Housing surfaces
If required
Cloth
Magnet and counterpiece
If required
Cloth
Reflector
If required
Cloth
7.2 Cleaning
The unit may only be
cleaned when it is
cool, since there is
burn hazard. Do not
use any cleaning solutions. The
heater must not be touched, as
hand sweat considerably
shortens the life cycle of the
unit.
What
Part
Frequency
Check all plug-in connections for correct fit.
Power plug
Weekly
Check the switch-off mechanism by opening the curing chamber during
Interlock
Weekly
program operation. If the fan continues to work, the safety switch
(interlock) is defective. In this case, contact your local Ivoclar Vivadent
Service Center.
Check the infrared heater (26) for damage of the glass.
Infrared heater (26)
Before the first
operation. Weekly
Check temperature and light performance
Test Set
Every 6 months
This chapter describes the user maintenance
and cleaning procedures. All other tasks
must be performed by qualified service
personnel at a certified Ivoclar Vivadent
Service Center.
7.1 Monitoring and maintenance
The time for these maintenance procedures
depends on the frequency of use and the
working habits of the users. For that
reason, the recommended times are only
approximates.
Disconnect power before maintenance
and cleaning, since there is a risk of
electrical shock.
The following parts have to be cleaned from time to time:
Use only original Ivoclar Vivadent dust
filters.