p19
Reassembly
Pump on pump support:
1
Affix the connection tube (V) to the other side of the
pump, as well as the hose connection to the inlet or
outlet of the vacuum system at the valve head (O).
2
Slip connecting tube (V) onto hose connection of
valve head.
3
Slide on the tube and the hose clip (U) until touching
the nose at the valve head.
4
Close hose clip (U) with flat pliers.
5
Put head cover cowling on.
6
Slide the head cover cowling in the grooves of the
caps (I) and under the connection fasteners (Q).
7
Install the washers. Use a Torx driver T20 to attach
the 4 screws holding the head cover cowling.
8
Tighten the fillister head screws (L) of the connection
fasteners (Q) with a Torx driver T20.
9
Close the hinged covers.
Pump at pump support:
Assemble catchpots with joint clips.
Replace diaphragms and valves of the opposite side of
the pump in the same way!
Step 1
Step 5
Step 8
Step 10
I
Q
- IEF100 (65 pages)
- IEF100 (65 pages)
- SE900 (37 pages)
- SE260 (31 pages)
- SE260 (31 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE410 (51 pages)
- SE410 (50 pages)
- SE640 (51 pages)
- SE640 (50 pages)
- SE275 (21 pages)
- SE245 (13 pages)
- SE675 (27 pages)
- SG100 (9 pages)
- SG500 (12 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- HE33 (26 pages)
- HE33 (26 pages)
- HE99X (26 pages)
- HE99X (26 pages)
- HE-PLUS System (26 pages)
- HE-PLUS System (26 pages)
- SUB Series (31 pages)
- SUB Series (31 pages)
- TE70X (31 pages)
- TE70X (31 pages)
- PR648 (11 pages)
- PR150 (9 pages)
- PR645 Immunoblot XL (21 pages)
- TE22 (35 pages)
- TE22 (35 pages)
- TE62 (35 pages)
- TE22 (31 pages)
- DQ300 (30 pages)
- DQ300 (30 pages)
- UVIS20 (21 pages)
- UVIS20 (21 pages)
- SP-2001 Vision Life Spectrophotometer (66 pages)
- Single Channel Variable Pipette (8 pages)
- PR250 (18 pages)
- PP24 (10 pages)