beautypg.com

Specifications – Hoefer VP200 User Manual

Page 11

background image

p3

2. Specifications

Max. pumping speed 50/60 Hz

2.0/2.3 m

3

/h

Max. pumping speed 60 Hz

1.2/1.4 cfm

Ultimate (total) pressure

7/5 mbar/torr

(absolute)

Ultimate (total) pressure

12/9 mbar/torr

(absolute) with gas ballast

Max. permitted outlet pressure 1.1 bar
(absolute)

Motor power

180 W

Motor protection

overload cutout
(manual reset)

Rated current draw
120 V~ 60 Hz

1

3.0 A

230 V~ 50/60 Hz

2

1.4/1.6 A

Degree of protection IEC 529

IP 40

Operating temperature

10–40 °C

Storage temperature

-10–60 °C

Operating relative humidity

30–85%

(non-condensing)

A-weighted emission

45 dB(A)

sound pressure level,
uncertainty KpA, 3 dB(A)

Rated motor speed at 50/60 Hz 1500/1800 rpm

Inlet/outlet port hose nozzle

10 mm

tubing size

Dimensions (w × h × d)

319 × 243 × 309 mm

Weight

13.6 kg

Product certifications

CE 89/336/EEC
(EMC directive)
CE 72/23/EEC
(LV directive)
EN-61010-1 (IEC 1010-1)
UL 3101-1, CSAA22.2
1010-1)

1

Supplied voltage must be within +5, -10% of the pump rating.

2

Supplied voltage must be within ±10% of the pump rating.

This declaration of conformity
is only valid for the instrument
when it is:

• used in laboratory locations.

• used as delivered from Hoefer,

Inc. except for alterations
described in the user manual.

• connected to other CE labeled

instruments or products
recommended or approved by
Hoefer, Inc.