Troubleshooting – Hoefer SE1200 User Manual
Page 22

•
p13
4. Troubleshooting
problem
solution
Gel or cellophane cracks
All gels
Handle the gel and cellophane with care. Fully
hydrate the cellophane with distilled water just before
use and avoid stretching it during assembly.
If the cellophane is drying too quickly, instead of
hydrating it in distilled water, soak it in the same
glycerol solution used for the gel.
Always remove all air bubbles before drying.
Lower the heat setting.
Load frame assemblies into the dryer from the top
down. Avoid dripping onto a partially dried gel.
Gradient gels or gels
Dehydrate the gel in 35% ethanol and 2% glycerol
≥0.75 mm thick with
before drying. If gels still crack, increase the
acrylamide ≥10% T
percentage of glycerol up to 5%.
Note: With a higher percentage of glycerol, a longer
drying period is required. The gels may also be flex-
ible and feel tacky when dry.
Fluor-treated gel
Use cellophane instead of mylar.
Use a thinner gel (≤0.75 mm).
- IEF100 (65 pages)
- IEF100 (65 pages)
- SE900 (37 pages)
- SE260 (31 pages)
- SE260 (31 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE410 (51 pages)
- SE410 (50 pages)
- SE640 (51 pages)
- SE640 (50 pages)
- SE275 (21 pages)
- SE245 (13 pages)
- SE675 (27 pages)
- SG100 (9 pages)
- SG500 (12 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- HE33 (26 pages)
- HE33 (26 pages)
- HE99X (26 pages)
- HE99X (26 pages)
- HE-PLUS System (26 pages)
- HE-PLUS System (26 pages)
- SUB Series (31 pages)
- SUB Series (31 pages)
- TE70X (31 pages)
- TE70X (31 pages)
- PR648 (11 pages)
- PR150 (9 pages)
- PR645 Immunoblot XL (21 pages)
- TE22 (35 pages)
- TE22 (35 pages)
- TE62 (35 pages)
- TE22 (31 pages)
- DQ300 (30 pages)
- DQ300 (30 pages)
- UVIS20 (21 pages)
- UVIS20 (21 pages)
- SP-2001 Vision Life Spectrophotometer (66 pages)
- Single Channel Variable Pipette (8 pages)
- PR250 (18 pages)
- PP24 (10 pages)