1
Install the safety lid
The cassettes and electrode panels are color coded to
match the leads in the lid: Orient the lid so that the
grey half of the cassettes and the grey anode panel
face the anode (+), or red lead, and the black half of
the cassettes and the black cathode panel face the
cathode (–), or black lead.
2
Use only an approved power supply such as the
Hoefer PS2A200, PS200HC, or PS300B. Make sure
the power supply is off and all controls are set to
zero. Plug the red lead into the red output jack and
the black lead into the black output jack. In most
systems, the red lead is the anode (+), and the black
lead is the cathode (–).
3
Set the power supply
Constant current mode is recommended. If constant
voltage mode is selected, carefully monitor the current
(increased current increases Joule heating). If the
current exceeds 1 A, decrease the voltage. If available,
set the power supply timer for no more than
two hours.
Note: The two red caps on the lid
accommodate the banana plugs on
the SE600 model immersible heat
exchanger (irrespective of
the orientation).