After the separation, Quick, high-voltage runs, Slower, lower voltage runs – Hoefer HE33 User Manual
Page 16

•
p8
Quick, high-voltage runs
Certain applications, such as screening samples or
checking sample purity, can be accomplished quickly
under high voltage conditions. Chill the base (-20 °C)
and limit the run to 5 minutes or less at 500 V.
Slower, lower voltage runs
A voltage gradient of 12 V/cm (150 V) separates 0.1
to 23 kb fragments of a Hind III digest of λ DNA
in 30 to 40 minutes (using 1% agarose gel and
0.5X TBE running buffer). Alternatively, using the
same solutions, this sample could be run at 24 V/
cm (300 V) with acceptable band resolution in 20 to
30 minutes. Chill the base before use.
Table 1: Voltage settings and recommended
run settings
†
voltage
gradient
time
(V)
(V/cm)
(min)
500
40
5*
400
31
10*
300
24
20*
200
16
30 to 40
150
12
30 to 60
* For rapid runs of 20 minutes or less, use 0.5X TBE and chill
the base to -20 °C before use.
†
Voltage and times are for 1% Agarose NA, 0.5X TBE and a
chilled base.
After the separation
Important! Turn off the power supply, disconnect the
leads, and remove the lid.
If no ethidium bromide was added to the gel or
sample before the run, stain the gel in a solution of
0.5 to 1.0 µg/ml ethidium bromide in water or buffer.
Clean the unit as described below.
Note: To calculate the voltage
gradient, divide the voltage
setting by the distance between
the electrodes (12.7 cm).