Fluke Biomedical 1060AM User Manual
Page 7

Introduction
Specifications
1
1-3
Digital Inputs
16 digital inputs with internal pull-up resistors
1.2.2 Analog Specifications
High Voltage
Regulated 500 to 2500 VDC, < 1 mV ripple, digitally controlled with 1 V resolution
Max
500 microamps at 1400 V
Input Circuitry:
Pulse mode input range 3 mV to 7 V suitable for BF3 detectors, proportional
counters, Geiger Mueller tubes, and photo multiplier tubes used with scintillators.
Polarity selection: (PMT input only)
High Disc., 0 - 7.50 VDC; set via software
Low Disc., 0 - 1.230 VDC; set via software
Jam detection (anti-jam)
1.2.3 Miscellaneous Specifications
Power Requirements Approximately 12 VDC @ 200 mA - Nominal Voltage range is 7 VDC (minimum)
to 16 VDC (maximum). With special adapter can accommodate 24 VDC.
Enclosures
Two standard enclosures are available (all dimensions in inches unless other-
wise indicated):
An acrylic plastic tube housing for indoor applications.
P/N
Outside Dimensions
Mounting Hole Pattern
1060AM-CL-xx 3.5 dia x 10
-
(8.89 x 25 cm)
A plastic rectangular housing for NEMA 4 applications (outdoor or indoor).
P/N
Outside Dimensions
Mounting Hole Pattern
1060-NM-xx 3.8 x 11.125 x 4.0
2.9375 x 9.375
(9.65 x 28.26 x 10.16 cm)
(7.46 x 23.81 cm)
User Interface
Via any IBM compatible Personal Computer
Model 1060 Circuit
Board Size
Approximate size: 2.8 in x 8.775 in (7.11 x 22.29 cm) with mounting holes on
2.4 in x 8.375 in (6.1 x 21.27 cm) centers
Auxiliary Voltage
An auxiliary voltage of +/- 12 VDC @ 50 mA, non-inductive load, is available to
provide power to preamplifiers.
Temperature/Humidity
Range
The humidity range is 5% to 95% rh, non-condensing over the entire temperature
range of 32° to 122°F (0° to 50°C).
Shock and Vibration
Mechanical shock and vibration specifications are per ANSI N42.17A, section 8.4
and 8.5.
Software Specifications
Language
"C" The software written in Intel "iC96", Version X243
Operating System
Real-time, interrupt driven, embedded system