Calibration, maintenance and troubleshooting, 1 calibration, 2 maintenance – Fluke Biomedical 1060AM User Manual
Page 23

Calibration, Maintenance and Troubleshooting
Calibration
5
5-1
CAUTION
CAUTION
CAUTION
Section 5
Calibration, Maintenance and Troubleshooting
5.1 Calibration
5.1.1 Precautions
Only trained and qualified personnel operating within the protocols of an approved radiation protection 
program should be permitted to calibrate any radiation detector. 
5.1.2
Exposure
The Model 1060 is calibrated at the factory against NIST traceable standards. The calibration is detector 
specific. 
Calibrations other than to standard nuclides are available upon request. Model 1060s will always provide 
raw counts, but if calibrated, will also provide engineering unit information. 
5.2 Maintenance
The instrument power should be turned OFF any time that the instrument is opened to prevent possible 
damage. 
 
 
 
Many components on the printed circuit board are 
static sensitive. ESD precautions should be 
observed when handling the printed circuit board 
assembly. 
 
 
 
If readings from a device's position indicates that 
the radiation rate is relatively high, CAUTION 
should be observed when replacing or working on 
this unit. High rates can quickly cause the 
operator's cumulative exposure to increase with the 
potential for injury. 
 
 
 
There is a potential shock hazard when the Model 
1060 board is exposed and the unit is operating. Be 
especially careful not to touch any components in 
the high voltage area, or connections to the 
detector. 
 
