Operation
Terminal Operation
3
3-3
NOTE
RRS
r
Read mR/Hr (latest second mR/Hr value)
RSH
r
Read Sieverts (averaged over 60 minutes)
RSM
r
Read Sieverts (averaged over 60 seconds)
RSS
r
Read Sieverts (latest second Sievert value)
RTA
Read TAU (deadtime)
SCx
s
Set Conversion Factor x, where x = 0 to 9
*SCR
s
Set Conversion Factor CPM to mR/Hr (enter CPM per mR/hr)
SDO
Set Digital Output (Sets 16 digital outputs)
SFD
s
Set output Format Decimal (integer)
SFE
s
Set output Format Engineering (default, x.xxExx)
SFH
s
Set output Format Hex (integer, 16 hex digits)
SGT
s
Set GM Threshold (Sets Thresh., stores in EEPROM)
SHD
s
Set High Discriminator (Sets Hi Discrim., stores in EEPROM)
SHS
s
Set High Scale analog output (entered in units of mR/hr)
SHV
s
Set High Voltage (Sets HV, stores setting in EEPROM)
SIC
s
Set Input Control
SID
s
Set unit Identification
SLD
s
Set Low Discriminator (Sets Lo Discrim., stores in EEPROM)
SLS
s
Set Low High Scale analog output (enter in units of mR/hr)
*STA
s
Set TAU (deadtime correction);
VER
Read firmware Version and Revision level
XSM
s
Exit Setup Mode
XTM
t
Exit Test Mode
Commands SCR and STA must be used to enter
the CPM to mR/hr conversion factor and deadtime
correction. These values must be entered
whenever a new detector is installed. These values
can be found on the detector calibration sheet.
3.3 LED Definitions
1.
Heartbeat (flash pulse indicates proper operation)
2.
HV ON
3.
RS-232 (Port 1) Command accepted and processed
4.
RS-232 (Port 1) Character received
5.
RS-485 (Port 2) Command accepted and processed
6.
RS-485 (Port 2) Character received
7.
RS-485 (Port 3) Command accepted and processed
8.
RS-485 (Port 3) Character received
3.4 DIP Switch Definitions
1.
ON = HV ON
2.
ON = Port 1 (RS-232) Setup commands available
3.
ON = Port 2 (RS-485) Setup commands available
4.
ON = Port 3 (RS-485) Setup commands available
5.
Open
6.
Open