Operation, 1 communication protocol, 2 terminal operation – Fluke Biomedical 1060AM User Manual
Page 17

Operation
Communication Protocol
3
3-1
Section 3
Operation
3.1 Communication Protocol
Communication Port Command Entry Syntax
RS-232 Communication Port 1:
This port (9-pin connector) is used for local communication.
Commands from the computer are of the form:
CMD
Responses from the 1060 are of the form:
RSP
where: CMD = TLA (Three Letter Acronym) command
RSP = Response from 1060
Carriage
Return
RS-485 Communication Port 2 and Port 3:
These ports are used in multi-drop configuration; commands must be preceded with the 1060 unit ID:
ID
where ID=1060 ID,
Responses from the 1060 with matching ID are of the form:
where ID=1060 ID,
3.2 Terminal Operation
Command Set
The following COMMAND SET is comprised of STANDARD, SETUP and RADIATION commands.
STANDARD commands are always available to all communication ports. SETUP commands (marked "s")
are used to configure the 1060 and are made available through Switch S2 as follows:
Port 1 (RS-232), S2, Position 2 ON = Setup commands available.
Port 2 (RS-485), S2, Position 3 ON = Setup commands available.
Port 3 (RS-485), S2, Position 4 ON = Setup commands available.
RADIATION commands (marked "r") are used to poll the 1060 for specific radiation data.
To enter Setup mode, enter the command ESM (Switch 2 must be ON). This will enable the SETUP and
disable the RADIATION commands. To exit Setup mode, enter XSM. This will disable the SETUP and
enable the RADIATION commands.