Patient conditions – Fluke Biomedical BP Pump 2 User Manual
Page 51

Operation
Simulations
3
3-15
Standard BP
Preset # 1
C
u
ff: External
OPTIO
N
S
CUFF
ready ....
120/
8
0 (93)
8
0 BPM
0.6
8
cc
fas07.eps
2. Press the
OPTIONS
soft key to scroll through the simulation choices.
Parameters for Standard BP simulations are shown in Table 3-2.
Table 3-2. Standard Blood Pressure Simulations
Simulation
Number
Blood Pressure
(mmHg) (MAP)
Heart Rate (bpm)
Pulse volume
(cc)
1 120/80
(93) 80
0.68
2 150/100
(116) 80
0.65
3 200/150
(166) 80
0.60
4 255/195
(215) 80
0.55
5 60/30
(40) 80
0.75
6 80/50
(60) 80
0.71
7 100/65
(76) 80
0.69
3. Press the
CUFF
soft key to select Internal Adult or External cuff.
4. Press
Start
on the NIBP unit.
5. Compare values on the NIBP unit with those on the Tester.
6. Press
HOME
to return to the
Main
menu.
Patient Conditions
The Patient Condition simulations are intended to provide some basic patient
variations.
To perform Patient Conditions simulations, do the following:
1. Access these simulations by pressing the
5 PATIENT CONDITIONS
key.
The
Patient Condition
screen appears.