Pressure generation, Electrical ecg, Heart rate for nibp simulations – Fluke Biomedical BP Pump 2 User Manual
Page 23

Introduction and Specifications
Specifications
1
1-11
Range ..................................................................... 0 mmHg to +400 mmHg
Resolution............................................................... 0.1 kPa
1 mmHg
1 cmH
2
O
1 inH
2
O
0.1 psi
Resolution (High Accuracy Version) ....................... 0.01 kPa
0.1 mmHg
0.1 cmH
2
O
0.1 inH
2
O
0.01 psi
Accuracy (Standard Version)
0 to 300 mmHg ................................................... ±0.5 % of reading ±1 mmHg
301 to 400 mmHg .............................................. ±2 % of reading
High-Accuracy Version ........................................... <0.8 mmHg (0.1 kPa)
Pressure Generation
Pressure Generator, Static Pressure Range .......... 50 mmHg to +400 mmHg
Difference between target pressure
and actual pressure ................................................ ±10 mmHg from 100-400 mmHg with
a minimum volume of 300 cc
Internal Leak Rate .................................................. <2 mmHg per minute, with a minimum
volume of 300 cc
Electrical ECG
Signals .................................................................... RA, LA, RL, LL, V
Waveform ............................................................... Lead II
Amplitude................................................................ 1 mV peak (±10%)
Connections............................................................ Signals available via the optional ECG
adapter
Heart Rate for NIBP Simulations
Heart Rate Accuracy
With ECG disabled ............................................ ±1 BPM
With ECG enabled .............................................. ±1 BPM
Except for the following Patient Conditions:
Weak Pulse, Tachycardia, Obese, Geriatric....... ±1% ±1 BPM
Patient Condition Mild Exercise .......................... ±1.5% ±1 BPM
Patient Condition Strenuous Exercise ................ ±3% ±1 BPM