Fluke Biomedical 942A-200L-M4 User Manual
Page 8

942A-200L-M4 & 942A-200L-M5 UDR 
Operator Manual 
1-4
Power ON/OFF
Alternate action pushbutton
Relay Outputs
HIGH Alarm – 2 sets. SPDT rated 5 A @ 120 VAC (one set 120 Vac 
powered for use with optional local alarm) 
Auxiliary HIGH Alarm – 2 sets. SPDT rated 5 A @ 120 VAC (uses RATE 
relay) 
WARN Alarm – 2 sets. SPDT rated 5 A @ 120 VAC 
RATE Alarm – Not Implemented 
FAIL Alarm – 2sets. SPDT rated 5 A @ 120 VAC 
High Voltage
Variable from 450 to 1500 VDC @ 0.2 mA
Detector Power
± 15 v @ 20 mA
Recorder/Computer 
Outputs 
Two 4 to 20 mA (500
Ω load max. each), and one 0 to 10 VDC (1 KΩ load
min.), scaled to the ratemeter range 
For the 942A-200L-M4 only, two additional analog outputs are provided 
from the Model 942-200-95-M1 Dual analog Output Option board. The 
ranges are 0-10mV and 0-5 VDC (1Kohm load min.). For further 
information, refer to the 942-200-95D Analog Output Option Instruction 
Manual, provided separately. 
Remote Alarm Ack 
Input 
Optically isolated DC input
Detector Input
Digital pulse train. Detector may be located up to 1500 feet from the 
ratemeter. Individually adjusted upper and lower level discriminators limit 
MeV response to desired range. 
Statistical Accuracy
Statistical accuracy is user selectable at 2%, 5%, 10%, or fixed at a one 
second update. Count time automatically adjusts per selected accuracy 
and count rate (not recommended for safety related applications). 
UDR Electronic 
Accuracy 
± 1% of the displayed value (± 1 digit)
Single Channel 
Analyzer 
For the 942A-200L-M4 only, a Model 942-200-90-M1 Single Channel 
Analyzer option circuit board is provided. This board provided the ability to 
perform pulse height analysism by counting the only pulses that fall 
between two adjustable voltage windows. For further information, refer to 
the 942-100-90D Single Channel Analyzer Instruction Manual, Provided 
separately. 
Dimensions 
(H x W x D) 
3.5 in. x 5.6 in. x 13.5 in. 
(8.9 cm x 14.2 cm x 34.3 cm) 
Weight
Approximately 5 lbs. (2.3 kg)
Power
120/240 VAC
± 10 %, 50/60 Hz, 28 watts
Operating Temperature
32
°F to 122 °F
0
°C to + 50 °C
Relative Humidity
0 to 99 % non-condensing
Compatible Detectors
VICTOREEN 943-XXL Series Scintillation Detectors w/LED
Modification 
Information 
For the Model 942A-200L-M4, refer to Modification Instruction 942A-200L-
M4MS, provided in Appendix E 
For the Model 942A-200L-M5, refer to Modification Instruction 942A-200L-
M5MS, provided in Appendix E 
 
 
 
 
 