Table 1-3 set point functions, Switch position, Position – Fluke Biomedical 942A-200L-M4 User Manual
Page 10: Jumper

942A-200L-M4 & 942A-200L-M5 UDR
Operator Manual
1-6
Table 1-3 Set Point Functions
Switch Position
Setpoint
Units
Factory Setting
0
High Alarm Limit
CPM
1.00E5
1
Warn Alarm Limit
CPM
1.00E3
2
Resolving Time (Tau)
Minutes/count
0.00E0
3
Full Scale Value
CPM
1.00E7
4 Overrange
CPM
1.00E7
5 Conversion
Constant
-
1.00E0
6 Background
Subtract CPM
0.00E0
7
Low Scale Value
CPM
1.00E1
8
Calibration Mode
Seconds
6.0E1
9
Not Used
N/A
E0007 *
A thru E
Not Used
N/A
E0007 *
F Unit
I.D.
N/A
1.00E0
*= Press ENTER push button to clear error code.
Table 1-4 Standard Jumper Placement
Position
Jumper Description
OUT JP1 Reset
2-3
JP2
Standard 27256 PROM
IN
JP3-1
Fixed 1 Second Update
IN
JP3-2
Fixed 1 Second Update
IN
JP3-3
Latches alarms until acknowledged
IN
JP3-4
Gives fail condition without counts, in five minutes
IN
JP3-5
Alarms do not respond to check source counts
1-2
JP4
Negative Input Pulse
1-2
JP5
Negative Shield Polarity
1-2 JP6
Anti-Jam
Active
1-2
JP7
Anti-Jam scintillation type detector