Position, Jumper, Connect – Fluke Biomedical 942A-200L-M4 User Manual
Page 69

Appendix
Connector Designations
A
A-5
Table A-9. Standard Jumper Placement
Position
Jumper Description
OUT JP1 Reset
2-3
JP2
Standard 27256 PROM
IN
JP3-1
Fixed 1 Second Update
IN
JP3-2
Fixed 1 Second Update
IN
JP3-3
Latches alarms until acknowledged
IN
JP3-4
Gives fail condition without counts, in five minutes
IN
JP3-5
Alarms do not respond to check source counts
1-2
JP4
Negative Input Pulse
1-2
JP5
Negative Shield Polarity
1-2 JP6
Anti-Jam
Active
1-2
JP7
Anti-Jam scintillation type detector
Table A-10. UDR Jumper Setting Options
Jumper
Connect Description
JP1
-
Momentary connect resets microprocessor
JP2
1-2
PROM type 27128
JP2
2-3
PROM type 27256
JP3-1 IN/OUT
Accuracy
selection
JP3-2 IN/OUT
Accuracy
selection
JP3-3
IN/OUT
Alarm reset control, manual/automatic
JP3-4
IN/OUT
Low count fail control, enable/inhibit
JP3-5
IN/OUT
Alarm when check source is on, inhibit/enable
JP4
1-2
Negative input pulse selection
JP4
2-3
Positive input pulse selection
JP5
1-2
Positive shield (for negative pulse)
JP5
2-3
Negative shield (for positive pulse)
JP6
1-2
Anti-Jam fuse enable – Normal Operation
JP6
2-3
Anti-Jam fuse disable – Test Only
JP7
1-2
Scintillator detector anti-jam input
JP7
2-3
GM tube detector anti-jam