Jumper, Function, Position – Fluke Biomedical 942A-200L-M4 User Manual
Page 54: Operation, 9 jumper selectable options

942A-200L-M4 & 942A-200L-M5 UDR
Operator Manual
3-14
3.9 Jumper Selectable Options
NOTE
The unit must be turned off when changing jumpers.
Table 3-1 Model 942A-200L-M4 and 942A-200L-M5 Option Jumpers (Refer to drawing 942-200-10,
Appendix A, for Jumper Location)
Jumper
Function
Position
Operation
JP1 Microprocessor
Reset
Out
IN
Normal Operation (factory) Momentary
connect to reset microprocessor without
cycling AC power Not Applicable
JP2 PROM
Type
1-2
2-3
PROM 27128
PROM 27256 (factory)
JP3-
1/JP3-2
Statistical Accuracy
(It should be noted that this
accuracy refers only to the
statistical interpretation of
detector counts, and not
detector accuracy)
Out/Out
Out/In
In/Out
In/In
2% Accuracy – 9604
5% Accuracy – 1537 Target Counts
10% Accuracy – 384 Target Counts
Fixed one second display update (factory)
JP3-3 Alarm
Acknowledge
In
Out
Manual Acknowledge (factory)
Automatic
JP3-4
Fail Alarm
In
Out
Enable No Counts Fail Alarm (factory)
Inhibit No Counts Fail Alarm
JP3-5
Check Source Alarm
In
Out
Alarm Inhibited (factory)
Alarm Enabled
JP4
Input Pulse Selection
1-2
2-3
- Input Pulse (Scintillation Detectors)
(factory)
+ Input Pulse (GM Detectors)
JP5
Shield Polarity Selection
1-2
2-3
+ Shield for Negative Pulse (factory)
- Shield for Positive Pulse (GM)
JP6
Anti-Jam Fuse Selection
1-2
2-3
Out
Enable for Normal Operation (factory)
Anti-Jam Circuit fuse bypassed (testing)
Anti-Jam Circuit Disabled (for testing only)
JP7
Detector Type for Anti-Jam
Circuit
1-2
2-3
Scintillation (factory)
GM Tube