Gas standard – Fluke Biomedical VT305 User Manual
Page 66

VT305
Users Manual
52
Gas Standard
Gas Standard
Temperature
Pressure
Ambient Temperature and Pressure (ATP)
Current Gas temperature
Current ambient pressure
Ambient Temperature and Pressure Dry (ATPD)
Current Gas temperature
Current ambient pressure
Ambient Temperature and Pressure Saturated (ATPS)
Current Gas temperature
Current ambient pressure
Ambient pressure at 21
°
C 21.0
°
C (70
°
F) Current
ambient
pressure
Standard Conditions USA (STP)
21.0
°
C (70
°
F)
1013.25 mbar (760 mmHg)
Standard Conditions USA Humid (STPH)
21.0
°
C (70
°
F)
1013.25 mbar (760 mmHg)
Body Temperature and Pressure Saturated (BTPS)
37
°
C (99
°
F) Current
ambient
pressure
Body Temperature and Pressure Dry (BTPD)
37
°
C (99
°
F) Current
ambient
pressure
Standard Conditions DIN1343 (0/1013)
0
°
C (32
°
F)
1013.25 mbar (760 mmHg)
Standard Conditions ISO 1-1975 (DIN 102 (20/981))
20
°
C (68
°
F)
981 mbar (736 mmHg)
API Standard Conditions (15/1013)
15
°
C (60
°
F)
1013.25 mbar (14.7 psia)
Cummings Standard (25/991)
25
°
C (77
°
F)
991 mbar (500ft Höhe)
20
°
C / 1013 mbar (20/1013)
20
°
C (68
°
F)
1013.25 mbar (760 mmHg)