Vt305, Trigger, Standard – Fluke Biomedical VT305 User Manual
Page 25

Gas Flow Analyzer
Operation
11
individual settings. Table 5 is a list of screens that show in
the display.
Table 5. Settings Screens
Screen Description
VT305
Owner:
Company:
Next Calib:
Last Calib:
Software:
Hardware:
Information
Shows device data. You can
set Owner and Company data
fields with the browser-based
configurator. See the Product
Configuration section.
Battery
Shows the current charge of
the battery.
Table 5. Settings Screens (cont.)
Screen Description
Ethernet
Default
IP: 192.168.1.1
Subnet: 255.255.255..0
Ethernet
The Ethernet screen is used to
set the Ethernet communication
parameters.
Trigger
Adult
Start: 60ms
Flow: >3.0
l/min
End: 60ms
Flow: >3.0
l/min
Set Trigger
The Trigger events screen is
used to set when the Product
calculates volume, and
respiratory parameters. The
factory defaults show adult,
pediatric, and high frequency
trigger configuration. See the
Measure Key Respiratory Data
section.
Standard
ATP
Amb. Temperature/Pressure
Set Gas Standard
The Product calculates the
measured flow and volume
values for the set standard. See
the Gas Standard after the
Specifications section.