Setup for ventilator measurements – Fluke Biomedical VT305 User Manual
Page 30

VT305
Users Manual
16
Setup for Ventilator Measurements
To test and calibrate ventilators, use the inlet pipe between
the breathing circuit and the Product, as shown in
Figure 13. Use the filter to improve laminarity of the flow.
This improves measurement accuracy.
Breathing circuit
Inlet Pipe
Fluke VT305
Testlung
Positive flow direction
gyo052.eps
Figure 13. Ventilator Connections