Detailed Filter Positioning

AP Full Spine 20-25 cm (Very Large Patient)(72”FFD)

5

5-5

Figure 5-6

5.4 AP Full Spine Over 25 cm (Very Large Patient) (72" FFD)

Breast Shields (57-409)

72" AP/PA Wedge Filter (57-405)

Build-Up Filter (Thick) (57-434)

Build-Up Filter (Thin) (57-430)

Gonad Shield (57-408) or (57-444)

Position the top edge of the AP/PA Filter at the level of the external auditory meatus. Attach the Breast
Shields on top of the AP/PA Wedge Filter. Place the Build-Up Filter (thick) on top of the Breast Shields,
with the top edge at the base of the nose. Position the Build-Up Filter (thin) on top of the Thick Build-Up
Filter to provide additional filtration for the odontoid process (see Figure 5-7). An increase in technique is
necessary as the patient measurement increases. The measurement is taken in the AP position at the
level of the third lumbar vertebra.


Figure 5-7


Thin Build-Up Filter
(57-430)


Thick Build-Up Filter
(57-434)

AP/PA Filter
(57-405)

Breast Shield
(57-409)

Gonad Shield
(57-408) or (57-444)