Detailed Filter Positioning
AP Full Spine 20-25 cm (Very Large Patient)(72”FFD)
5
5-5
Figure 5-6
5.4 AP Full Spine Over 25 cm (Very Large Patient) (72" FFD)
Breast Shields (57-409)
72" AP/PA Wedge Filter (57-405)
Build-Up Filter (Thick) (57-434)
Build-Up Filter (Thin) (57-430)
Gonad Shield (57-408) or (57-444)
Position the top edge of the AP/PA Filter at the level of the external auditory meatus. Attach the Breast
Shields on top of the AP/PA Wedge Filter. Place the Build-Up Filter (thick) on top of the Breast Shields,
with the top edge at the base of the nose. Position the Build-Up Filter (thin) on top of the Thick Build-Up
Filter to provide additional filtration for the odontoid process (see Figure 5-7). An increase in technique is
necessary as the patient measurement increases. The measurement is taken in the AP position at the
level of the third lumbar vertebra.
Figure 5-7
Thin Build-Up Filter
(57-430)
Thick Build-Up Filter
(57-434)
AP/PA Filter
(57-405)
Breast Shield
(57-409)
Gonad Shield
(57-408) or (57-444)