Fluke Biomedical 6000D Impulse Getting Started User Manual
Page 31
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
19
Noise Immunity
Wave ................................................................... Sine
Line Frequency....................................................50 or 60 Hz (± 0.5 Hz)
Amplitude ............................................................0.0 to 10.0 (by 0.5) mV
Accuracy..............................................................± 5%
Transvenous Pacer Pulse Simulation
Widths
Range..............................................................0.1, 0.2, 0.5, 1.0, and 2.0 ms
Accuracy..........................................................±5 % of setting
Amplitudes...........................................................0 (off) and ±2, ±4, ±6, ±8, ±10, ±12, ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, and
±700 mV
Accuracy..............................................................± (10 % of setting + 0.2 mV)
Arrhythmia Selections
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)
Demand...........................................................30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA
Supraventricular
Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia