Impulse 6000D, 7000DP
Getting Started Manual
16
Sample rate............................................................250 kHz (4
μs sample)
Maximum Average Power.....................................12 W, equivalent to 10 defib pulses of 360 J every 5 minutes
Oscilloscope Output
Autorange............................................................2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output .............................................................BNC
Output impedance...........................................50
Ω (nominal)
Delay...............................................................50 ms (nominal)
Accuracy .........................................................
±5 % of nominal
Charge Time Measurement
Range..................................................................0.1 to 100.0 s
Accuracy .............................................................
±0.05 s, typical
Synchronization Test (Elective Cardioversion )
Delay Time Measurement
Timing window ................................................ECG R-wave peak to the defib pulse peak
Range..............................................................-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms
following the R-wave peak.
Resolution .......................................................1 ms
Accuracy .........................................................
±1 ms
ECG waves
Normal Sinus Rhythm (NSR) ..........................30 to 180 (by 1) BPM
Atrial fibrillation................................................Coarse and fine
Monomorphic Ventricular Tachycardia ...........120 to 240 (by 5) BPM
Asystole...........................................................Flat line