Defibrillator analyzer specifications, Energy output measurement – Fluke Biomedical 6000D Impulse Getting Started User Manual
Page 27
Defibrillator/Transcutaneous Pacer Analyzer
Specifications
15
Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Autoranged Measurement ....................................0.1 to 600 J
Accuracy
0.1 to 360 J .....................................................
±(1 % of reading + 0.1 J)
360 to 600 J ....................................................
±(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is
±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance...........................................................50
Ω
Accuracy..........................................................
±1 %, non-inductive (<2 μH)
Variable external load resistance (optional) .......25, 75,100, 125, 150, or 175
Ω, All values ±1 %, non-inductive (<2 μH)
Pulse trigger level .................................................20 V
Pulse width
Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................
±0.1 ms
Voltage
Range..................................................................20 to 5000 V
Accuracy..............................................................
±(1 % of reading + 2 V)
Current
Range..................................................................0.4 to 100.0 A
Accuracy..............................................................
±(1 % of reading + 0.1 A)