Defibrillator/Transcutaneous Pacer Analyzer

Specifications

15

Defibrillator Analyzer Specifications

Energy Output Measurement

Compatible Defibrillator Waveshapes.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Autoranged Measurement ....................................0.1 to 600 J
Accuracy

0.1 to 360 J .....................................................

±(1 % of reading + 0.1 J)

360 to 600 J ....................................................

±(1 % of reading + 0.1 J), typical

Note

For Pulsed Bi-Phasic defibrillator, specified accuracy is

±(1.5 % of reading + 0.3 J) on both ranges.

Load resistance

Resistance...........................................................50

Ω

Accuracy..........................................................

±1 %, non-inductive (<2 μH)

Variable external load resistance (optional) .......25, 75,100, 125, 150, or 175

Ω, All values ±1 %, non-inductive (<2 μH)

Pulse trigger level .................................................20 V
Pulse width

Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................

±0.1 ms

Voltage

Range..................................................................20 to 5000 V
Accuracy..............................................................

±(1 % of reading + 2 V)

Current

Range..................................................................0.4 to 100.0 A
Accuracy..............................................................

±(1 % of reading + 0.1 A)