Ecg waves – Fluke Biomedical 6000D Impulse Getting Started User Manual
Page 29

Defibrillator/Transcutaneous Pacer Analyzer
Specifications
17
Automated Defibrillator Test ECG Waves
Normal Sinus.......................................................30 to 300 (by 1) BPM
Ventricular Fibrillation..........................................Coarse and fine
Monomorphic Ventricular Tachycardia................120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia..................5 types
Asystole...............................................................Flat line
ECG Waves
ECG General
Lead configuration...............................................12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ......................................1000 Ω (nominal)
Rate accuracy .....................................................
±1 % of nominal
ECG Amplitudes
Reference lead ....................................................Lead 1
Settings ...............................................................0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Accuracy..............................................................
±2 % of setting, lead I and 2 Hz square wave
For performance waves and R-wave detection, other leads are proportional to Lead I in percentage per:
Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50
Leads V1 through V6 ......................................100
For normal sinus waves, other leads are proportional to Lead I in percentage per:
Lead I ..............................................................100
Lead II .............................................................150
Lead III ............................................................50