Fluke Biomedical 6000D Impulse Supplement User Manual
Page 8

Impulse 6000D/7000DP Users
Manual Supplement
4/08 7
Accuracy ......................................................... ±(10 % of setting + 0.2 mV)
Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent)
Lead II reference
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
67 %
100 %
33 %
67 %
67 %
67 %
67 %
67 %
67 %
Lead I reference
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
100 %
150 %
50 %
100 %
100 %
100 %
100 %
100 % 100 %
Arrhythmia Selections
Pacer Interactive (Transcutaneous pacer, Impulse 7000DP only)
Demand .......................................................... 30 to 360 (by 1) BPM
Asynchronous
Non-Capture
Non-Function
Threshold (Interactive pacing simulation only) 10 to 250 (by 10) mA
Supraventricular
Atrial Fibrillation Coarse
Atrial Fibrillation fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia