Fluke Biomedical 6000D Impulse Supplement User Manual
Page 6

Impulse 6000D/7000DP Users
Manual Supplement
4/08 5
Normal Sinus waves
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
70 %
100 %
30 %
24 %
48 %
100 %
120 %
112 %
80 %
Lead I reference
Performance waves and R wave detection
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
100 %
150 %
50 %
100 %
100 %
100 %
100 %
100 %
100 %
Normal Sinus waves
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
100 %
150 %
50 %
24 %
48 %
100 %
120 %
112 %
80 %
ECG Normal Sinus
Rates ................................................................... 10 to 360 (by 1) BPM
ECG High Level Output (BNC Jack)
Amplitude
Range ............................................................. 0.5 V per mV of reference lead setting
Accuracy .........................................................
5 %
Output Impedance .............................................. 50 Ω
ECG on Defibrillator Input Load
Same as the LEAD II amplitude but limited to ±4 mV
ECG Performance Waves
Square wave ....................................................... 2.0 and 0.125 Hz
Triangular wave .................................................. 2.0 and 2.5 Hz