Fluke Biomedical 6000D Impulse Supplement User Manual
Page 4

Impulse 6000D/7000DP Users
Manual Supplement
4/08 3
Scope Output
Autorange ........................................................... 2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output ............................................................. BNC
Output impedance ........................................... 50
Amplitude Accuracy ........................................
5 %
Charge Time Measurement
Range ................................................................. 0.1 to 100.0 s
Accuracy .............................................................
0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement
Timing window ................................................ ECG R-wave peak to the defib pulse peak
Range ............................................................. -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380
ms following the R-wave peak.
Resolution ....................................................... 1 ms
Accuracy .........................................................
1 ms
ECG waves
Normal Sinus Rhythm (NSR) .......................... 10 to 180 (by 1) BPM
Atrial fibrillation ................................................ Coarse and fine
Monomorphic Ventricular Tachycardia ........... 120 to 240 (by 5) BPM
Asystole .......................................................... Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus ...................................................... 10 to 300 (by 1) BPM