Fluke Biomedical 6000D Impulse Supplement User Manual
Page 3

Manual Supplement
Impulse 6000D/7000DP Users
2
4/08
Accuracy .............................................................
0.1 ms
Voltage
Range ................................................................. 20 to 5000 V
Accuracy .............................................................
(1 % of reading + 2 V)
Current
Range ................................................................. 0.4 to 100.0 A
Accuracy .............................................................
(1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range ................................................................. 1 % to 99 %
Accuracy .............................................................
1 digit
Interphase delay (biphasic and pulsed biphasic)
Range ................................................................. 0.1 ms to 9.9 ms
Accuracy .............................................................
0.1 ms
Frequency (pulsed biphasic only)
Range ................................................................. 2000 Hz to 8000 Hz
Accuracy .............................................................
1 % of reading
Duty cycle (pulse biphasic only)
Range ................................................................. 1 % to 99 %
Accuracy .............................................................
1 digit
Sample rate ........................................................... 250 kHz (4
s sample)
Maximum Average Power .................................... 12 W, equivalent to 10 defib pulses of 360 J every 5 minutes