Change #3, Change #4, Change #5 – Fluke Biomedical 6000D Impulse Supplement User Manual
Page 11

Manual Supplement
Impulse 6000D/7000DP Users
10
4/13
Change #3,
48038
On page 38, under Pacemaker Input, change:
From:
Accuracy …………..±1 % non-inductive (<2
H)
To:
Accuracy……………±2 % non-inductive (<2
H)
Change #4
On page 3 add the following Caution:
Caution
To avoid damage to the product or defibrillator under test, do not use adhesive pads on the Analyzer. For
defibrillators with metal paddles, use Fluke Biomedical electrode adapters.
Change #5,
63282, 65427
On page 2, delete CAT I from the Symbols table and add:
Conforms to relevant South Korean EMC Standards
.
This manual is related to the following products: