Commissioning procedure – Flowserve OR 52-5 User Manual
Page 28
28
Commissioning Procedure
continued
Indicating decimal point
The 7-segment display can show a decimal point.
This setting does not have any effect on the adjusted limit
values and time delays.
Cut off power supply!
Open housing lid of the measuring transducer and set code switch 7 on lid plate to ON.
Setting actual value output
The actual value output can be changed from 0–20 mA to 4–20 mA.
Cut off power supply!
Open the housing lid and remove wire link 2 from the lid plate.
Wire link 2 set: actual value output 0–20 mA
Wire link 2 removed: actual value output 4–20 mA
ppm
sec
cal
0
OIL
Decimal point
Danger
The terminal strips of the measuring transducer are live during operation.
This presents the risk of severe injuries due to electric shock. Cut off
power supply before opening the housing lid.