beautypg.com

3 general instructions, Center adapter plate to shaft figure 5 – Flowserve SLM-6000 User Manual

Page 4

background image

4

3.1 Ensure that the bearing assembly is mounted correctly (central and square) to the

pump pedestal and securely fastened.

3.2 Ensure that all fit locations on the pedestal are clean and free of burrs.

3.3 Clean the pump shaft/sleeve surface, making sure it is clean of product and rust

and has no burrs.

3.4 If the pump design requires it, bolt seal adapter plate to equipment case/pedestal

(hand tighten only).

3.5 Locate the seal fit of the adapter to the shaft by use of a dial indicator or centering jig.

The equipment shaft must be concentric to the seal adapter bore to within 0.38 mm

(0.015"). See Figure 5.

2.6 Ensure that the shaft or shaft sleeve has a 1.6 mm (0.06") x 30° chamfer on the

leading edge to prevent O-ring damage at seal installation.

2.7 Hardened shaft sleeves must not be used when cartridge sleeve locking is

accomplished by set screws bearing directly into the shaft sleeve. Any of the sleeve

collars that clamp the cartridge sleeve to the shaft sleeve may be used on hardened

shaft sleeves.

3 General Instructions

Center Adapter Plate to Shaft

Figure 5

Centering Jig

Adapter to be Concentric with

Shaft within 0.38 mm (0.015")

3.6 For adapters provided with a centering jig, simply insert the centering jig into the

adapter and tighten adapter bolts alternately, adjusting the cover location until the

centering jig can be removed by hand without binding.

3.7 Apply a light coat of O-ring lubricant to the seal sleeve O-ring. (Silicone grease must

be used with EPDM gaskets.)

Warning: Do not apply anti-seize or other lubrication to the pump shaft/sleeve. Keep

the shaft/sleeve clean and dry. The use of lubricants will cause improper clamping

pressure by the clamp collar.

This manual is related to the following products: