14 

 

2.

Press “Mode” key until “pH” icon and “AUTO” icon light 
up. 

3.

Rinse the pH electrode probe with distilled water and 
immerse in the sample to be measured. Remove any air 
bubbles trapped around the probe by shaking or stirring 
the probe. 

4.

Press the “Mea.” key. The “WAIT” icon will start flashing. 
The unit is waiting for a stable reading. The display will 
track the pH value as sensed by the pH electrode probe. 

5.

When the “WAIT” icon disappears, the reading is then 
“HOLD” and will not respond to further changes from 
sample. The pH value shown is the pH value of the sample 
at the set sample temperature. 

[Note: For samples that are inherently unstable, the instrument will 
not AUTOLOCK. In this case, use the pH NON‐ AUTOLOCK mode for 
measurements.] 

Measurement with an ATC/Temp probe in the pH 

NON‐AUTOLOCK mode 

1.

Connect the pH electrode to the BNC connector and the 
ATC/Temp probe to the ATC/Temp connector of the unit. 
The “ATC” icon will light up. 

2.

Press “Mode” key until “pH” icon light up. 

3.

Rinse the pH electrode and ATC/temp probe with distilled 
water and immerse in the sample to be measured. 

4.

Allow sufficient time for the display to stabilize. The 
instrument will display the pH value of the sample at the 
displayed sample temperature. 

Measurement with manual temperature 
compensation in the pH NON‐AUTOLOCK mode 

1.

Connect the pH electrode to the BNC connector of the 
unit. The “MAN” icon will light up. 

2.

Press “Mode” key until “pH” icon light up. 

3.

Rinse the pH electrode probe with distilled water and 
immerse in the sample to be measured. 

4.

Allow sufficient time for the display to stabilize. The 
instrument will display the pH value of the sample at the 
set sample temperature. 

 

Display and Key Functions 

A. Display 

 

Figure 2 

1. WAIT  
This will be displayed when 
the unit is still waiting for a 
stable reading or end point 
sensing. 

6. HOLD 
This will indicate that the reading 
is frozen during AUTOLOCK mode. 

2. BAT‐ 
Low Battery Indicator 

7. EFF (%)‐ 
This will be displayed if the user is 
viewing the efficiency of the 
electrode. It is recommended to 
use a new electrode when the 
efficiency value is less than 75%. 

3. ATC/MAN 
ATC indicator will be 
displayed 
if a temperature probe is 
connected otherwise the 
MAN indicator will be 
displayed. 

8. pH/mV 
Unit and mode indicators. 

4. STAND/SLOPE 
This indicator will flash if 
the STAND or SLOPE is not 
yet calibrated. This indicator 
will remain lit‐up if the 
STAND or SLOPE has been 
calibrated respectively. 

9. Main display for pH, mV 
and probe efficiency values 

5. AUTO‐ 
Auto Lock mode indicator 

10. Temperature and unit display