beautypg.com

Ph calibration buffer table, Waterproof – Electro-Chemical Devices (ECD) H10 User Manual

Page 5

background image

16 

 

pH Calibration Buffer Table 

The temperature coefficient of pH calibration buffers 4.01, 6.86, 
7.00,9.18 and 10.01 are stored inside the instrument. The buffers 
used to calibrate the instrument must exhibit the same temperature 
characteristics as the stored values. 
 
Temperature coefficient of the pH buffers 

°C 

4.00 

6.86 

9.18 

4.01 

7.00 

10.01 

4.01 

6.98 

9.46 

4.01 

7.11 

10.32 

4.00 

6.95 

9.39 

4.01 

7.08 

10.25 

10 

4.00 

6.92 

9.33 

4.00 

7.06 

10.18 

15 

4.00 

6.90 

9.28 

4.00 

7.03 

10.12 

20 

4.00 

6.88 

9.23 

4.00 

7.01 

10.06 

25 

4.00 

6.86 

9.18 

4.01 

7.00 

10.01 

30 

4.01 

6.85 

9.14 

4.01 

6.98 

9.97 

35 

4.02 

6.84 

9.10 

4.02 

6.98 

9.93 

40 

4.03 

6.84 

9.07 

4.03 

6.97 

9.89 

45 

4.04 

6.83 

9.04 

4.04 

6.97 

9.86 

50 

4.06 

6.83 

9.02 

4.06 

6.97 

9.83 

55 

4.07 

6.83 

8.99 

4.08 

6.97 

9.80 

60 

4.09 

6.84 

8.97 

4.10 

6.98 

9.78 

 

 

Waterproof 

Though the Model H10 meter is housed in a watertight case, DO 
NOT 
use it underwater. The watertight case prevents permanent 
damage to the unit if accidentally dropped into non‐corrosive 
solutions.  
Follow these steps immediately if the unit is immersed in any 
solution: 

1.

Rinse unit carefully with distilled water. After rinsing and 
drying, inspect and clean connectors to remove all 
contaminants that may affect probe connections. 

2.

Wait for the unit and probe to dry completely before 
resuming operation. 

3.

 If the unit does not function correctly after steps1 and 2, call 
ECD for possible repair or replacement (see Warranty).