beautypg.com

General introduction, Initial inspection, Error displays and troubleshooting – Electro-Chemical Devices (ECD) H10 User Manual

Page 4

background image

 

General Introduction  

Thank you for selecting the Model H10 pH meter. The Model H10 is 
a precision tool that measures pH, mV and temperature. A built‐in 
microprocessor stores, calculates and compensates for all 
parameters related to pH determinations including pH electrode 
temperature characteristics, electrode slope deviations and buffer 
solutions. 
This meter has a waterproof IP67 case. The mechanical keys are 
highly reliable with tactile and audio feedback. This meter is 
powered by four AAA‐size alkaline batteries. The meter also displays 
a “BAT” message when the batteries are in need of replacement. Re‐
calibration is not required when power is restored.  
The front of the meter has a large LCD that displays pH or mV and 
temperature simultaneously along with user prompts and mode 
indicators. The unit prompts the user through calibration and 
measurement procedures.  
An AUTOLOCK feature for both pH and mV measurements enables 
the unit to automatically sense the end point and “HOLD” the 
display to indicate the end point value of a measurement. 
AUTOLOCK and user prompts help eliminate most errors in 
determining pH and mV values, resulting in precise, repeatable and 
error‐free measurements. The Model H10 can also be used in non‐
AUTOLOCK mode. 
The Model H10 is available with pH, ORP and ATC (Automatic 
Temperature Compensation) probes. Other features include 
electrode offset recognition, electrode slope recognition; electrode 
efficiency display, built‐in buffer coefficients, automatic or manual 
temperature compensation, long battery life, and 50/60Hz AC noise 
rejection. This meter is user‐friendly, for field, industrial and 
laboratory applications. 

Initial Inspection 

Carefully unpack the unit and accessories. Inspect for damages made 
in shipment. If any damage is found, notify your ECD representative 
immediately. All packing materials should be saved until satisfactory 
operation is confirmed. 

17 

 

Error Displays and troubleshooting 

Main 

Display 

Possible Causes 

Corrective 

Actions 

“Er1” 

● “Stand” was pressed before the 

electrode and ATC/Temp probe 
settled to within +/‐1.5 pH of 
the buffer value. 

●pH electrode offset is greater / 

less    than +/‐1.5 pH. 

●pH electrode is faulty.

 

●Press “Clear” key, allow

sufficient time for the 
electrode and ATC/Temp 
probe to stabilize, re‐
press “Stand” key to start 
the calibration 
procedure. 

●Replace the buffer and 

/or the pH electrode. 
Press “Clear” key to 
recalibrate meter. 

●Replace electrode. 

“Er2” 

● “Slope” was pressed before 

the electrode and ATC/Temp 
probe settled to within 30% of 
the buffer value. 

● Buffer 4.00, 4.01, 9.18 and 

10.01 is not correct. 

●pH electrode slope is off by 

more than 30% of ideal slope. 

●Allow sufficient time for 

the electrode and 
ATC/Temp probe to 
stabilize, re‐press “Slope” 
key to continue the 
calibration procedure. 

●Check if the correct 

buffer is used. 

●Replace the buffer and 

/or the pH electrode. 
Press “Clear” key to 
recalibrate meter. 

“Er3” 

Temperature is out of the 
0.0 to 60.0°C range.

Bring the buffer 
temperature within range.

“Over” 

Or 

“Undr” 

●Measured pH is out of the 2.00 

to 16.00 pH range. 

●Measured mV is out of the 1999 

to 1999 mV range.  

●Measured temperature is out of 

the –10.0 to 120.0°C range. 

●Bring sample pH into the

correct measuring range.

●Bring sample ORP into 

the correct measuring 
range. 

●Bring sample 

temperature into the

correct measuring range.