StockWeigh 500

D3632 US-B

12

Short Form Calibration

The Short Form Setup & Calibration procedure allows you to change the “SETUP” and “CAL”

numbers of the indicator.

Setup Number

The “SETUP” numbers affects some parameters that also appear in the Long Form

Calibration. Following is a list of functions that are controlled by the “SETUP” number:
Weigh Method (W MTHD)

Display Units (LB-KG)

Display Counts (COUNT)

Gain

Scale Capacity

Calibration Number

The “CAL” number is adjusted to make the scale read the proper weight for different load

cells and to make accuracy adjustments on a scale system.
If a StockWeigh 500 indicator is provided with StockWeigh load cells as a package, the

calibration number is adjusted at the factory for maximum accuracy. Systems that are

assembled from separate components are given “STANDARD SETUP AND CALIBRATION

NUMBERS” and are not factory calibrated. These systems should be checked and adjusted

if necessary using known weights to insure accuracy. See “Calibrating the Scale for

Maximum Accuracy” below.
Following is a list of Standard Setup and Calibration numbers for StockWeigh 3300, 6600

and 14000 load cells. If you are not using StockWeigh load cells, contact Digi-Star Tech

Support at 920-563-9700 for setup and calibration information:

Scale Type

Unit

Setup #

Calibration #

StockWeigh 3300

Lb

413004

776

StockWeigh 6600

Lb

413007

1560

StockWeigh 10000

Lb

443010

14040

StockWeigh 14000

Lb

443015

13828

StockWeigh 3300

Kg

812002

352

StockWeigh 6600

Kg

812003

708

StockWeigh 10000

Kg

843005

6368

StockWeigh 14000

Kg

843007

6272