beautypg.com

Hydrating the ph probe, Cleaning the ph probe, 0 cleaning the bluelab ph probe – Bluelab pH Controller Connect Manual User Manual

Page 17: 0 hydrating the ph probe, English

background image

English

17

26.0 Cleaning the Bluelab pH Probe

To ensure accurate readings the pH probe tip needs to be rinsed in water and cleaned

prior to calibration using the following instructions. After cleaning, use the probe

straight away, or place the storage cap on the probe tip. Always ensure the cap

contains enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.

Remove storage cap from pH probe.
Hold the top of the storage cap, twist the cap to

loosen then remove.

Rinse pH probe tip under fresh tap water.

Fill a small plastic container with clean tap

water. Add a small amount of Bluelab pH Probe

Cleaner or mild detergent (dishwashing liquid).

Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the pH probe

on the side of the container as this may cause

damage to the probe. Rinse well under fresh

running water to remove all traces of the

detergent mixture.
If the probe tip requires removal of heavy

contamination: Gently brush around the

glassware with a few drops of Bluelab pH Probe

Cleaner or mild detergent (dishwashing liquid)

and a soft toothbrush.

Rinse well under fresh running tap water to

remove all traces of the detergent mixture.

Calibrate pH probe after cleaning, see section

14.0. After calibration use straight away or

store pH probe in the storage cap, ensuring KCl

Storage Solution covers the probe tip.

1

3

2

4

5

6

7

6

5

3

1

4

2

7

25.0 Hydrating the pH probe

Hydrate the pH probe in Bluelab pH Probe KCl Storage Solution when:
• the probe tip has not always been stored in KCl storage solution, to improve the

reading response speed.

• the probe tip has been accidentally allowed to dry out.

Never store the pH probe in RO (Reverse Osmosis), De-ionized or Distilled water. Pure water changes the chemistry
in the reference, causing the probe to die.

Clean the pH probe tip. Ensure the probe tip is cleaned

before hydrating. See section 26.0 for instructions.
Add enough Bluelab pH Probe KCl Storage Solution to a

plastic container to submerge the pH probe tip.

Loosen, then remove the storage cap (if required). Place

the pH probe upright in a the KCl solution.

Leave to soak for up to 24 hours. After hydration, always

calibrate the pH probe to ensure accuracy, see section 14.0.

Bluelab pH
Probe KCl
Storage
Solution

1

3

2

4

This manual is related to the following products: