0 ph calibration cont, English – Bluelab pH Controller Connect Manual User Manual
Page 11

English
11
14.0 pH calibration cont.
To calibrate the pH
Clean pH probe tip.
See section 26.0 (the pH probe does not
require cleaning before the fi rst use).
The calibration default is set at two point
calibration. If a three point calibration is
required:
a) Press ‘menu’.
b) Press ‘
’ to fi nd ‘2/3 point cal’. Press ‘enter’.
c) Press ‘
’ to select ‘3’. Press ‘enter’. When
‘saved’ is displayed on screen, three point
calibration is now available.
In separate plastic containers, prepare a
small amount of: fresh tap water, pH 7.0, pH
4.0 and/or pH 10.0 calibration solutions.
pH 7.0 calibration
a) Ensure the pH controller connect is plugged in.
b) Press and hold ‘cal’ for three seconds. ‘pH 7
Calibrate’ will be displayed.
c) Place the clean pH probe tip in the pH 7.0
calibration solution. Press ‘cal’.
d) Calibration is complete when all the ‘’s
on the screen become solid. The screen
will display ‘OK’ and the ‘pH 7’ indicator
will appear to indicate successful pH 7.0
calibration.
e) Now you can calibrate to pH 4.0 and/or pH
10.0
pH 4.0 and/or pH 10.0 calibration
a) Rinse the probe tip in fresh tap water,
shake off excess water. Place the clean
pH probe tip in either pH 4.0 or pH 10.0
calibration solution. Press ‘cal’.
b) Calibration is complete when all the ‘’s
on the screen become solid. The screen
will display ‘OK’ and the ‘pH 4’ or ‘pH 10’
indicator will appear to indicate successful pH
calibration.
c) If you require three point calibration, repeat
‘5a’ and ‘5b’ using the pH 4.0 or pH 10.0,
whichever solution was not used.
d) The pH controller connect is now
calibrated, and ready for use.
After pH calibration, the pH controller
connect reverts to ‘Monitor’ mode.
Change to ‘Control’ mode if required.
NOTE: If ‘Failed
calibration, see section 27.0 Troubleshooting.
1
2
3
4
5
6
press cal button
press cal button
successful
pH 7, 4 and pH 10
calibration
successful pH 7 calibration
ready for pH 4 or pH 10 calibration
pH 7.0
calibration
solution
pH 4.0 or
pH 10.0
calibration
solution
clean the pH probe
rinse in tap water
successful
pH 7 and pH 4
calibration