Electromagnetic compatibility information – Beurer BM 19 User Manual
Page 88

88
Electromagnetic Compatibility Information
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The BM 19 uses RF energy only for its internal function. Therefore, its RF emissi
ons are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The BM 19 is suitable for use in all establishments, including domestic establish
ments and those directly connected to the public lowvoltage power supply net
work that supplies buildings used for domestic purposes.
Harmonic emissions IEC 6100032
Not applicable
Voltage fluctuations/flicker emissions
IEC 6100033
Not applicable
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 6100042
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Power frequency (50/60 Hz)
magnetic field IEC 6100048 3 A/m
3 A/m
Power frequency magnetic fields should be at levels characte
ristic of a typical location in a typical commercial or hospital
environment.