Specifications – Beurer BM 19 User Manual
Page 21
21
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement (in addition to
, the cardiac arrhythmia icon
appears in the dis
play),
4. the cuff tube is not inserted properly,
5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Reinsert the batteries if necessary, or else
replace them.
9. Specifications
Model no.
BM 19
Measurement
method
Oscillometric, noninvasive blood
pressure measurement on the upper
arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 – 199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac
curacy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 147 mm x W 104 mm x H 66 mm
Weight
Approx. 310 g (without batteries)
Cuff size
23 to 33 cm
Permissible operating
conditions
+10 °C to +40 °C, ≤85 % relative air
humidity (noncondensing)
Permissible storage
conditions
20 °C to +70 °C, ≤85 % relative air
humidity, 700 –1060 hPa ambient
pressure
Power supply
4 x 1,5 V
AA batteries
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 4 x 1.5 V AA batter
ies, storage pouch
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 6060112
and is subject to particular precautions with regard to elec
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus
tomer Service address or found at the end of the instructions
for use.