beautypg.com

Specifications – Beurer BM 19 User Manual

Page 21

background image

21

1. inflation pressure is higher than 300 mmHg,

2. blood pressure is unusually high or low,

3. you move or talk during the measurement (in addition to

, the cardiac arrhythmia icon

appears in the dis­

play),

4. the cuff tube is not inserted properly,

5. the cuff is inflated for longer than 16 seconds.
In the above cases, you must repeat the measurement. Make

sure that the cuff tube is properly inserted and that you do

not move or talk. Re­insert the batteries if necessary, or else

replace them.

9. Specifications

Model no.

BM 19

Measurement

method

Oscillometric, non­invasive blood

pressure measurement on the upper

arm

Measurement range

Cuff pressure 0 – 300 mmHg,

systolic 40 – 280 mmHg,

diastolic 40 – 280 mmHg,

Pulse 40 – 199 beats/minute

Display accuracy

Systolic ± 3 mmHg,

diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement inac­

curacy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 147 mm x W 104 mm x H 66 mm

Weight

Approx. 310 g (without batteries)

Cuff size

23 to 33 cm

Permissible operating

conditions

+10 °C to +40 °C, ≤85 % relative air

humidity (non­condensing)

Permissible storage

conditions

­20 °C to +70 °C, ≤85 % relative air

humidity, 700 –1060 hPa ambient

pressure

Power supply

4 x 1,5 V

AA batteries

Battery life

For approx. 250 measurements,

depending on the blood pressure level

and/or pump pressure

Accessories

Instruction for use, 4 x 1.5 V AA batter­

ies, storage pouch

Classification

Internal supply, IPX0, no AP or APG,

continuous operation, type BF applied

part

Technical information is subject to change without notification

to allow for updates.

• This unit is in line with European Standard EN 60601­1­2

and is subject to particular precautions with regard to elec­

tromagnetic compatibility (EMC). Please note that portable

and mobile HF communication systems may interfere with

this unit. More details can be requested from the stated Cus­

tomer Service address or found at the end of the instructions

for use.