Monroe Electronics Electrostatic Fieldmeter - Static Monitor - model 177A User Manual
Page 49
48
177A reply to read coil:
Unit# command
# of data
Coil Value
crcH
crcL
1-255 0x01
1
0=Off, 0xff=On
Write Channel LEDs or Relays (write coil):
From Master (PC software or PLC):
Unit# Command
CoilAdrH
CoilAdrL
New Value
0
crcH
crcL
1-255 0x05
0
0-24
0=Off, 0xff=On
The request consisted of the Unit address and the command code 05 followed by a 2 byte starting
address, specifying the variable to be set, a data byte, specifying the binary value to set ( ON = 0xff,
OFF = 0 ) and an additional byte always set to 0.
177A reply to write single coil is identical to the received packet:
Unit# Command
CoilAdrH
CoilAdrL
New Value
0
crcH
crcL
1-255 0x05
0
0-24
0=Off, 0xff=On
Read Channel display (read register):
From PC or PLC:
Unit# Command
RegAdrH
RegAdrL
LenH LenL crcH crcL
1-255 0x03
0
0-3
0
1
RegAdrL is as follows:
0
Channel 1
1
Channel 2
2
Channel 3
3
Channel 4
177A reply:
Unit# Command
# of data
dataH dataL crcH crcL
1-255 0x03
1
##
##
Write Channel display (write register):
From PC or PLC:
Unit# Command
RegAdrH
RegAdrL
dataH dataL crcH crcL
1-255 0x06
0
0-3
##
##
177A reply to write single Register is identical to the received packet:
Unit# Command
RegAdrH
RegAdrL
dataH dataL crcH crcL
1-255 0x06
0
0-3
##
##
Note:
If the Unit# is 0, the 177A units will execute the Write-Coil and Write-Register commands, but No reply.