APPENDIX IV
Communication with ME 177A using MODBUS Protocol
Read Channel LEDs or Relays (read coil)
Write Channel LEDs or Relays (write coil)
Read Channel display (read register)
Write Channel display (write register)
Read Channel LEDs or Relays (read-coil):
From Master (PC software or PLC):
Unit# Command CoilAdrH CoilAdrL lenHigh lenLow crcH
crcL
1-255 0x01 0
0-24
0
1
The request consisted of the unit# and the command code 01 followed by a 2 byte starting address,
specifying the variable to be read.
Length high must be 0, length low is 1.
CoilAdrH is 0
CoilAdrL is as follows:
0
Channel 1 Alarm (red) LED
1
Channel 2 Alarm (red) LED
2
Channel 3 Alarm (red) LED
3
Channel 4 Alarm (red) LED
4
Channel 1 Warning (Amber) LED
5
Channel 2 Warning (Amber) LED
6
Channel 3 Warning (Amber) LED
7
Channel 4 Warning (Amber) LED
8
Channel 1 OK (Green) LED
9
Channel 2 OK (Green) LED
10
Channel 3 OK (Green) LED
11
Channel 4 OK (Green) LED
12
Channel 1 Alarm Relay
13
Channel 2 Alarm Relay
14
Channel 3 Alarm Relay
15
Channel 4 Alarm Relay
16
Channel 1 Warning Relay
17
Channel 2 Warning Relay
18
Channel 3 Warning Relay
19
Channel 4 Warning Relay
20
Channel 1 OK Relay
21
Channel 2 OK Relay
22
Channel 3 OK Relay
23
Channel 4 OK Relay
24
System Alarm Relay