Casella CEL Sensus data logger unit User Manual
Page 37

11.
POWER SUPPLIES
11.1 Solar Power
The Sensus Logger incorporates a solar
power regulator as standard. This
regulator is designed to operate in
conjunction with a nominal 12 V (30 W)
panel with a short circuit current of no
more than 2 A.
The solar panel should be wired
to the connector identified in Figure 25.
The pin assignments for the
solar panel connector are as follows.
Pin
Assignment
1
Solar panel negative terminal,
2
Solar panel positive terminal.
11.2 Battery
The Sensus Logger incorporates a 12 V
lead-acid battery charger as standard.
This charger is designed to operate in
conjunction with a nominal 12 V lead-acid
battery and supplies a maximum charge
current of 0.5 A.
The lead-acid battery should be
wired to the connector identified in
Figure 26.
The pin assignments for the 12
V lead-acid battery connector are as
follows.
Pin
Assignment
1
Battery negative terminal,
2
Battery positive terminal.
01081
Pin 1
Pin 2
Figure 25: Solar power connector
Figure 26: Battery connector
01081
Pin 1
Pin 2
Power Supplies
SENSUS Digital Data Logger -
Users Handbook
Page 37 of 44
- CEL-62X (37 pages)
- CEL-63X (87 pages)
- CEL-430 (8 pages)
- CEL-440 (70 pages)
- CEL-480 (2 pages)
- CEL-231 (2 pages)
- CEL-283 (6 pages)
- CEL-228 (6 pages)
- CEL-275 (56 pages)
- CEL-393 (99 pages)
- CEL-193 (20 pages)
- CEL-224 (4 pages)
- Vortex personal sampling pumps (14 pages)
- Vortex Ultraflow area sampling pump (12 pages)
- Tuff personal sampling pumps (38 pages)
- Tuff personal sampling pump (1 page)
- Environmental enclosure for Microdust (10 pages)
- Sampling heads for personal dust sampling (24 pages)
- Apex personal sampling pumps (40 pages)
- Apex personal sampling pump (2 pages)
- CEL-630 (2 pages)
- AFC123 (24 pages)
- Airborne bacteria sampler (16 pages)
- CEL-620 (2 pages)
- CEL-500 (16 pages)
- CEL-400 (2 pages)
- CEL-450 (77 pages)
- CEL-272 (8 pages)
- CEL-420 (16 pages)
- CEL-320 (24 pages)
- CEL-300 (2 pages)
- CEL-268 (99 pages)
- CEL-350 (2 pages)
- CEL-35X dBadge (48 pages)
- CEL-315 (45 pages)
- CEL-180 (14 pages)
- CEL-281 (147 pages)
- Thermohygrograph standard model (12 pages)
- Tipping bucket rain gauge (16 pages)
- Thermohygrograph deluxe model (12 pages)
- Piche evaporimeter (4 pages)
- Nomad portable weather system (2 pages)
- Fortin barometer (11 pages)
- Campbell Stokes Sunshine recorder (2 pages)
- Barographs small and display pattern (2 pages)