Hot-air sterilization at 180°c – BINDER C 150 User Manual
Page 8

Hot-air sterilization at 180°C
Grenzert
Manual abort
of sterilization
3 sec.
Normal Display
Start the sterilization cycle
After the end of sterilization
Running sterilization
Turn off the unit,
or press “180°C”,
or open the outer door.
The unit is ready to operate.
Ineffective sterilization when aborting prematurely!
3 sec.
Do NOT open the doors until the
interior temperature has
dropped to 37°C. After a manual
abortion, the C 150 reverts to its
standard operational status. As
long as the interior temperature
remains above the temperature,
which was set to trigger the
temperature alarm, the C 150
will trigger alarm signals. You
can switch them off by pressing
“EXIT”. If necessary, repeat the
sterilization.
10 to 12 hours
Alternating display
Alternating display
showing the actual temperature
Copyright
BINDER GmbH 10/2012
Art. No. 7001-0235
©
Hot-air sterilization of the inner chamber
The hot-air sterilisation takes10 to
12 hours in 3 steps:
- Rapid heating-up
- Maintaining the inner
temperature for 4 hours,
on
- Cooling-down phase up to 37°C
which
ensures 180°C
all internal
surfaces for at least 30 minutes
Grenzert
GB
GB
Before starting the sterilization,
Empty the water
pan
pull out the
sensor without
any rotation from the connection
socket and remove it from the
inner chamber.
.
CO