Transilluminator Ballast Replacement
Overview
The purpose of this Technical Bulletin is to explain the process of uninstalling and reinstalling a
transilluminator ballast on a UVP transilluminator.
To complete this procedure, the following tools and parts will be required:
•
Phillips-head screwdriver
•
1/4” nut driver
•
New transilluminator ballast
Caution: This procedure requires a moderate level of technical competence. If you are not
comfortable working with electronics, tools and/or related components, contact UVP for
assistance.
Removal Procedure
When performing the following procedure, place all components (screws, nuts, etc.) in a secure
location as they will be reused for installation.
1. Turn off the transilluminator by placing the power switch on the front of the unit in the OFF
position. Then, unplug the unit from the wall power.
2. Remove the four Phillips-head screws and washers securing the transilluminator filter frame
to the body of the transilluminator. Lifting up, remove the transilluminator filter frame from the
transilluminator body. Place the filter frame aside.
Two of Four
Phillips-Head
Screws and
Washers
Transilluminator
Filter Frame
Transilluminator
Body