beautypg.com

UVP TB-107 Transilluminator Ballast User Manual

UVP Health and hygiene

background image

TB-107 Rev STD

T

T

e

e

c

c

h

h

n

n

i

i

c

c

a

a

l

l

B

B

u

u

l

l

l

l

e

e

t

t

i

i

n

n

T

T

B

B

-

-

1

1

0

0

7

7

Transilluminator Ballast Replacement

Overview

The purpose of this Technical Bulletin is to explain the process of uninstalling and reinstalling a

transilluminator ballast on a UVP transilluminator.

To complete this procedure, the following tools and parts will be required:

Phillips-head screwdriver

1/4” nut driver

New transilluminator ballast

Caution: This procedure requires a moderate level of technical competence. If you are not

comfortable working with electronics, tools and/or related components, contact UVP for

assistance.

Removal Procedure

When performing the following procedure, place all components (screws, nuts, etc.) in a secure

location as they will be reused for installation.

1. Turn off the transilluminator by placing the power switch on the front of the unit in the OFF

position. Then, unplug the unit from the wall power.

2. Remove the four Phillips-head screws and washers securing the transilluminator filter frame

to the body of the transilluminator. Lifting up, remove the transilluminator filter frame from the

transilluminator body. Place the filter frame aside.

Two of Four

Phillips-Head

Screws and

Washers

Transilluminator

Filter Frame

Transilluminator

Body