Test Products International 183A User Manual
Page 9

17
16
d. MEASURING AC AMPS
CAUTION!
Do not attempt to make a current measure
ment with the test leads connected in
parallel with the circuit to be tested. Test
leads must be connected in series with the
circuit.
WARNING!
Do not attempt to make a current measure
ment of circuits with more than 600V
present. Instrument damage and /or
personal injury may result.
Instrument set-up:
FUNCTION BLACK RED
MINIMUM
MAXIMUM
TEST LEAD
TEST LEAD
READING
READING
µA
COM
µAmA
0.1µA
4000µA
mA
COM
µAmA
0.01mA
400mA
10A
COM
A
0.001A
10.00A
Measurement Procedure:
1.
Disconnect power to the circuit to be measured.
2.
Plug the black test lead into the COM input jack.
3.
Plug the red test lead into the µAmA or A input jack
depending on the value of current to be measured..
4.
Set the rotary switch to the µA, mA or A function.
5.
Press the AC/DC pushbutton to set to AC mode.
6.
Connect test leads in series to circuit to be measured.
7.
Reconnect power to the circuit to be measured.
8.
Read the current on the 183A.
Optional Functions:
RANGE
REC
REL
HOLD
Manually select the appropriate range.
Activate MIN/MAX record mode (page 24).
Scroll between Hz, Duty, and Volts.
Activate REL mode (page 24).
Freeze the reading on the LCD.
Hz/DUTY
16
c. MEASURING DC AMPS
CAUTION!
Do not attempt to make a current measurement with
the test leads connected in parallel with circuit to be
tested. Test leads must be connected in series with
the circuit.
WARNING!
Do not attempt to make a current measurement of
circuits with more than 600V present. Instrument
damage and /or personal injury may result.
Instrument set-up:
FUNCTION BLACK RED
MINIMUM
MAXIMUM
TEST LEAD
TEST LEAD
READING
READING
40µA
COM
µAmA
0.01µA
40µA
µA
COM
µAmA
0.1µA
4000µA
mA
COM
µAmA
0.01mA
400mA
10A
COM
A
0.001A
10.00A
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Plug the black test lead into the COM input jack.
3.
Plug the red test lead into the µAmA or A input jack
depending on the value of current to be measured.
4.
Set the rotary switch to the 40µA, µA, mA, or A
function.
5.
Connect test leads in series to circuit to be measured.
6.
Reconnect power to the circuit to be measured.
7.
Read the current on the 183A.
Optional Functions:
RANGE
REC
REL
HOLD
Manually select the appropriate range.
Activate MIN/MAX record mode (page 24).
Activate REL mode (page 24).
Freeze the reading on the LCD.