Test Products International 183A User Manual

Page 9

background image

17

16

d. MEASURING AC AMPS

CAUTION!
Do not attempt to make a current measure
ment with the test leads connected in
parallel with the circuit to be tested. Test
leads must be connected in series with the
circuit.

WARNING!
Do not attempt to make a current measure
ment of circuits with more than 600V
present. Instrument damage and /or
personal injury may result.

Instrument set-up:

FUNCTION BLACK RED

MINIMUM

MAXIMUM

TEST LEAD

TEST LEAD

READING

READING

µA

COM

µAmA

0.1µA

4000µA

mA

COM

µAmA

0.01mA

400mA

10A

COM

A

0.001A

10.00A

Measurement Procedure:

1.

Disconnect power to the circuit to be measured.

2.

Plug the black test lead into the COM input jack.

3.

Plug the red test lead into the µAmA or A input jack
depending on the value of current to be measured..

4.

Set the rotary switch to the µA, mA or A function.

5.

Press the AC/DC pushbutton to set to AC mode.

6.

Connect test leads in series to circuit to be measured.

7.

Reconnect power to the circuit to be measured.

8.

Read the current on the 183A.

Optional Functions:

RANGE

REC

REL

HOLD

Manually select the appropriate range.

Activate MIN/MAX record mode (page 24).

Scroll between Hz, Duty, and Volts.

Activate REL mode (page 24).

Freeze the reading on the LCD.

Hz/DUTY

16

c. MEASURING DC AMPS

CAUTION!
Do not attempt to make a current measurement with
the test leads connected in parallel with circuit to be
tested. Test leads must be connected in series with
the circuit.

WARNING!
Do not attempt to make a current measurement of
circuits with more than 600V present. Instrument
damage and /or personal injury may result.

Instrument set-up:

FUNCTION BLACK RED

MINIMUM

MAXIMUM

TEST LEAD

TEST LEAD

READING

READING

40µA

COM

µAmA

0.01µA

40µA

µA

COM

µAmA

0.1µA

4000µA

mA

COM

µAmA

0.01mA

400mA

10A

COM

A

0.001A

10.00A

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug the black test lead into the COM input jack.

3.

Plug the red test lead into the µAmA or A input jack
depending on the value of current to be measured.

4.

Set the rotary switch to the 40µA, µA, mA, or A
function.

5.

Connect test leads in series to circuit to be measured.

6.

Reconnect power to the circuit to be measured.

7.

Read the current on the 183A.

Optional Functions:

RANGE

REC

REL

HOLD

Manually select the appropriate range.

Activate MIN/MAX record mode (page 24).

Activate REL mode (page 24).

Freeze the reading on the LCD.