Nor-Cal CDG 160 OP Lit User Manual
Page 24
![background image](/manuals/740022/24/background.png)
24
CDG160-OP-LIT
(2010-06)
Evacuate the gauge to a pressure according to the table
below:
F.S.
Recommended final pressure
for zero adjustment
1000 Torr
100 Torr
20 Torr
10 Torr
2 Torr
1 Torr
<5
×10
-2
Torr
<5
×10
-3
Torr
<1
×10
-3
Torr
<5
×10
-4
Torr
<1
×10
-4
Torr
<5
×10
-5
Torr
If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
the factory setting and adjust the zero again (
→ 30).
Operate the gauge for at least 2 hours under constant
ambient conditions (until the signal is stable).
Briefly press the
ø1.1 mm). The zero adjustment runs automatically. The
≤8 s)
is completed.
Press the button briefly
max. ø1.1 mm