Datamax I-4208 User Manual

Page 83

background image

I-Class

73

Advanced Entry Calibration (continued)

Step

Operator Action

Displayed Message

Comment

11

Proceed according to
your media type:

Die-cut stock: strip the
media from the backing
and then reinsert it into
the sensor; adjust the
Sensor Eye Mark over
the center of the backing.

Notched stock: adjust
the Sensor Eye Mark
over the center of the
notch.

Reflective stock: Adjust
the Sensor Eye Mark
over the center of the
facedown black mark.

Press the Key
to select the setting and
then record the sensor
reading in a table similar
to the one shown after
Step 12, as a TOF
Value.

Next, press the
Key to increment the
Gain Number.

Repeat for each of the
Gain Numbers (00-32).


HBJO!USBO!!!!!=zzz?

+11!!!!!!!!)1!.!42*

Or, for reflective media:

HBJO!SFGM!!!!!=zzz?

+11!!!!!!!!)1!.!42*

Where ‘yyy’ is a
numerical value
representing the
current sensor
reading: useful in
locating the best
sensor position.

; Note: Never
position the

sensor over a
perforation when
recording a
sensor reading.


; Note: Do NOT move the position of the Media Sensor after it has been

adjusted.